ACCU-CHEK Comfort Curve. Cat. No. 3149030; Ref. number 0314903001 with meter model number 2138930.
This recall has been terminated (originally issued August 5, 2006).
- Company
- Roche Diagnostics Corp.
- Recall Initiated
- June 22, 2006
- Posted
- August 5, 2006
- Terminated
- April 7, 2008
- Recall Number
- Z-1359-06
- Quantity
- 55,189 meters (for all Recalls Z-1354-1371-06)
- Firm Location
- Indianapolis, IN
- Official Source
- View on FDA website ↗
Reason for Recall
The meter gives an error message that can "actually" mean either a problem with the strip or a blood glucose too low to measure, but the meter error message only reports that there is a strip error.
Distribution
Worldwide, USA, Germany, Guam, North Mariana Islands, Panama and Virgin Islands.
Lot / Code Info
All units.
Root Cause
Other
Action Taken
A recall notification letter dated April 24, 2006 was sent to each customer alerting them to the fact that this error code has a dual meaning. New software will be installed in all meters when it becomes available.