Illumina MiSeqDx Cystic Fibrosis Clinical Sequencing Assay in vitro diagnostic system for use on the Illumina MiSeqDx Platform.
This recall has been terminated (originally issued December 23, 2014).
- Company
- Illumina, Inc.
- Recall Initiated
- November 6, 2014
- Posted
- December 23, 2014
- Terminated
- March 17, 2015
- Recall Number
- Z-0850-2015
- Quantity
- 19 units
- Firm Location
- San Diego, CA
- Official Source
- View on FDA website ↗
Reason for Recall
Illumina has confirmed a software limitation with MiSeq Reporter software (version 2.2.31) when using the MiSeqDx Cystic Fibrosis Clinical Sequencing Assay on the MiSeqDx instrument.
Distribution
US: MD, NY Non-US: France
Lot / Code Info
Instrument Serial No. M70106, M00533L, M70114. Lot No. Shipped: K103313, K103461, K103404, K103487.
Root Cause
Under Investigation by firm
Action Taken
A customer notification letter dated 11/12/14 was sent to all customers who purchased the MiSeqDx Cystic Fibrosis Clinical Sequencing Assay. The letter confirms that there is a software limitation with MiSeq Reporter software (version 2.2.31) when using the MiSeqDx Cystic Fibrosis Clinical Sequencing Assay on the MiSeqDx instrument. At certain genomic locations within the region sequenced by the assay, deletions cannot be reported by the software. The letter outlines the potential issues, the results of Illuminia's investigation, and the firm's next steps. Customers' with technical questions or concerns, are instructed to contact their local support team or Illumina Technical Support at techsupport@illumina.com. Customers with questions regarding the field action, are instructed to contact Illumina Regulatory Affairs at RA@illumina.com. Customers will be contacted by 11/13/14 by phone and a phone script was provided that includes the procedure, talking points, actions required, and next steps.