Reagent Kits/ containing flow cells, components of NextSeq 550Dx Instrument: NextSeq 550Dx High Output Reagent Kit v2.5 (75 cycles) IVD, REF: 20028870/NextSeq 550Dx HO FC Cart v2.5, 75 Cycles, REF: 20031098; NextSeq 550Dx High Output Reagent Kit v2.5 (300 cycles) IVD, REF: 20028871/NextSeq 550Dx HO FC Cart v2.5, 300 Cycles, REF: 20026365; CN NextSeq 550Dx High Output Reagent Kit v2.5 (75 cycles) IVD, REF: 20064341/CN NEXTSEQ 550Dx HO FLOW CELL V2.5, REF: 20062316; CN NextSeq 550Dx High Output Reagent Kit v2.5 (300 cycles) IVD, REF: 20064344/CN NEXTSEQ 550Dx HO FLOW CELL V2.5, REF: 20062316; CN NextSeq 550Dx Mid Output Reagent Kit v2.5 (150 cycles) IVD, 20064345/CN NEXTSEQ 550Dx MO FLOW CELL V2.5, REF: 20062530; CN NextSeq 550Dx Mid Output Reagent Kit v2.5 (300 cycles) IVD, 20064346/CN NEXTSEQ 550Dx MO FLOW CELL V2.5, REF: 20062530; /FLOWCELL, FIT FLUIDICS TESTING, REF: 15050205

This recall is currently active, issued December 4, 2025. It was issued by Illumina, Inc..

Company
Illumina, Inc.
Recall Initiated
October 28, 2025
Posted
December 4, 2025
Recall Number
Z-0884-2026
Quantity
11916
Firm Location
San Diego, CA
Official Source
View on FDA website ↗

Reason for Recall

An instrument intended for sequencing DNA libraries with in vitro diagnostic assays is used with reagent kits including flow cells that have flow cell gaskets that have a quality issue, which may lead to flow cell leaks during runs, which could cause run failures, that may result in: No/insufficient data output, instrument fluidics clogs, toxic substance exposure to user.

Distribution

Worldwide - US Nationwide distribution in the states of PA, NJ, CA, NC, UT, IL, MD, NY, VA, WV, SD, TX, AL, SC, TN, KY, MI and the countries of DK, SA, FR, DE, IT, CN, PL, GB, TH, GR, VN, BE, ES, IL, KR, AU, CH, BG, NL, IE, TW, TR, CZ, HU, NO, FI, JP, SG, LU, AT, AE, ZA, PT, LT, SE.

Lot / Code Info

NextSeq 550Dx UDI-DI: 00816270020125. Kit REF/Flow Cell REF/UDI-DI(Lot): 20028870/20031098/00816270020132(A184824-3, A185016-3, A183951-3, A182816-3, A183339-3, A182533-3, A183950-3, A184531-3, A182413-3, A184058-3, A183613-3, A183211-3, A183612-3, A183826-3, A182302-3, A183449-3, A182298-3, A182532-3, A182299-3, A183830-3, A183118-3, A182301-3, A184359-3, A184727-3); 20028871/20026365/00816270020118(A182717-3, A182051-3, A179804-3, A183343-3); 20064341/20062316/006975709330014(A183402-4, A183550-4); 20064344/20062316/006975709330038(A186185-4, A183170-4, A183404-4, A183555-4, A183557-4); 20064345/20062530/006975709330045(A186074-4, A183278-4, A183405-4, A184005-4); 20064346/20062530/006975709330052(A185054-4, A183279-4, A183409-4, A183568-4, A183569-4, A183570-4, A184776-4); 15050205(20935999, 20956354)

Root Cause

Vendor change control

Action Taken

On 10/28/2025, recall notices were emailed to customers who were asked to do the following: Dispose of reagents per local regulations. See Instrument Reference Guide: https://support.illumina.com/downloads/nextseq-550dx-instrument-reference-guide-1000000009513.html Illumina requests that this notice be communicated to all relevant personnel within your organization and to any external organization that has received the potentially affected reagent kits. If you experienced a run failure in a previous run using an affected flow cell, or experience leaking issues please contact Illumina Technical Support at techsupport@illumina.com Complete and return the verification form via email to techsupport@illumina.com