Mi1250 SYNCHRONY 2, Item Numbers: a) 39541, FLEXSOFT S-Vector; b) 39545, FLEX28 S-Vector; c) 39547, FLEX26 S-Vector; Cochlear Implant

This recall is currently active, issued July 12, 2024. It was issued by Med-El Elektromedizinische Gereate, Gmbh Furstenweg 77A Innsbruck Austria.

Company
Med-El Elektromedizinische Gereate, Gmbh Furstenweg 77A Innsbruck Austria
Recall Initiated
April 18, 2024
Posted
July 12, 2024
Recall Number
Z-2356-2024
Quantity
7 devices
Official Source
View on FDA website ↗

Reason for Recall

A small number of devices may have a higher probability of a crack within the internal electronics substrate which could result in a non-functional device.

Distribution

US, Durham, NC

Lot / Code Info

a) 39541, UDI/DI 09008737395414, Serial Numbers: 859106; b) 39545, UDI/DI 09008737395452, Serial Numbers: 860731, 860782, 863380; c) 39547, UDI/DI 09008737395476, Serial Numbers: 860757, 860772, 863386

Root Cause

Nonconforming Material/Component

Action Taken

MED-EL issued an Urgent-Medical Device Field Safety Notice (FSN) Medical Device Return notice to its sole US consignee (distributor) on 04/18/2024 via email. The notice explained the problem with the device and requested the return of all affected devices.