MED EL Cochlear Implant System, Mi1000 MED EL CONCERTO, sold under the following implant types: a. Mil000 CONCERTO PIN +FLEX28; b. Mi1000 CONCERTO +FLEX28; c. Mi1000 CONCERTO PIN +FLEXsoft; d. Mil000 CONCERTO +Standard; e. Mil000 CONCERTO +Medium; f. Mi1000 CONCERTO PIN +Standard; g. M11000 CONCERTO +Compressed Product Usage: The Mi1000 CONCERTO Implant is an implantable part of the MED EL Cochlear Implant System and can only be used together with compatible MED EL external components. The MED EL Cochlear Implant System is intended to evoke auditory sensations via electrical stimulation of the auditory pathways for severely to profoundly hearing impaired individuals who obtain little or no benefit from acoustic amplification in the best aided condition.

This recall has been terminated (originally issued October 16, 2018).

Company
Med-El Elektromedizinische Gereate, Gmbh Furstenweg 77A Innsbruck Austria
Recall Initiated
October 16, 2018
Terminated
May 22, 2020
Recall Number
Z-2136-2019
Quantity
7 devices
Official Source
View on FDA website ↗

Reason for Recall

Devices were distributed despite the Helium-fine leak test results not being within the cochlear implant standards specification.

Distribution

Worldwide distribution - US Nationwide in the states of California and Illinois. Countries of ARGENTINA, AUSTRALIA, AUSTRIA, BRAZIL, BULGARIA, CHINA, FRANCE, GEORGIA, KOREA, REPUBLIC OF, POLAND, RUSSIAN FEDERATION, SPAIN, SWEDEN.

Lot / Code Info

a. Mil000 CONCERTO PIN +FLEX28, Serial Number 510312;   b. Mi1000 CONCERTO +FLEX28, Serial Number 518214;   c. Mi1000 CONCERTO PIN +FLEXsoft, Serial Number 531793;   d. Mil000 CONCERTO +Standard, Serial Number 510079;   e. Mil000 CONCERTO +Medium, Serial Number 532046;   f. Mi1000 CONCERTO PIN +Standard, Serial Number 535616;   g. M11000 CONCERTO +Compressed, Serial Number 538434

Root Cause

Employee error

Action Taken

Med-EL initially notified customers who had not yet implanted the affected devices on about 10/16/2018 via telephone call and followed up with an "Urgent-Medical Device Field Safety Notice (FSN)." The notice instructed customers to not implant the device if still available in inventory. Customers who had implanted the device were notified on about 11/27/2018 via telephone call and followed up with an "Urgent-Medical Device Field Safety Notice (FSN)." The notice instructed customers to immediately inform affected clinics of the recall and the long term risk. MED-EL does not recommend a device explantation, as the device may work as intended for years before the electronics may become affected by humidity ingress. Ultimately, accumulation of internal moisture may cause complete failure of the device. Should any uncomfortable or painful sensation be experienced, please stop immediately using the device and contact MED-EL or your local support center for immediate support. MED-EL recommends regular implant checks at least once a year to check the implant status. Additionally the customers were requested to respond to the FSN and confirm receipt.