Centricity PACS-IW with Universal Viewer version 5.0. Used to display medical images (Including mammograms) and data from various imaging sources.

Company
Ge Medical Systems, Llc
Recall Initiated
October 28, 2022
Posted
December 16, 2022
Recall Number
Z-0721-2023
Quantity
1 unit
Firm Location
Waukesha, WI

Reason for Recall

There is the potential that after installation of the IAS tool, some studies can remain hidden without user notification. Access to those hidden studies requires GE Healthcare assistance, which can result in a delay in reading the study.

Distribution

Nationwide distribution to CA, CO, DE, FL, IA, IL, MA, MI, MN, MO, MS, MT, ND, NJ, NY, OK, OR, SD, TX, WI, WV, WY. International distribution to Austria, Brazil, France, Germany, Ghana, Ireland, Italy, Kuwait, Poland, Saudi Arabia, South Africa, Spain, Switzerland, United Kingdom.

Lot / Code Info

Installed Product ID: YP3150-Centricity Universal Viewer-01639018

Root Cause

Software Manufacturing/Software Deployment

Action Taken

GE Healthcare notified consignees on about 10/28/2022 via letter titled, "URGENT MEDICAL DEVICE CORRECTION - UPDATE." Consignees were instructed that the device can continued being used by following these instructions: 1) When a study is not present on the Study List, contact a GE Healthcare Service Representative for assistance in recovering the study. 2) In situations where study interpretation is time sensitive, interpret the study at the scanning unit or plan for an alternate means of reading studies for emergent cases. GE Healthcare will contact consignees to arrange for the correction. Consignees were also instructed to complete and return the acknowledgement form.