LOGIQ P10 R4.5 HD ultrasound system with system software version R4.5.7., Model Number 5877535
- Company
- Ge Medical Systems, Llc
- Recall Initiated
- September 18, 2025
- Posted
- October 15, 2025
- Recall Number
- Z-0156-2026
- Quantity
- 31 units
- Firm Location
- Waukesha, WI
Reason for Recall
The Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially lead to inappropriate clinical decisions impacting overall care.
Distribution
US Nationwide. Global Distribution.
Lot / Code Info
Model Number 5877535 UDI-DI: 00195278464668, Serial Numbers: LPZ490338 LPZ490347 LPZ490339 LPZ490147 LPZ490292 LPZ490019 LPZ490252 LPZ490348 LPZ490303 LPZ490304 LPZ490305 LPZ490306 LPZ490340 LPZ490350 LPZ490341 LPZ440658 LPZ440673 LPZ440674 LPZ490337 LPZ440641 LPZ440192 LPZ440258 LPZ440694 LPZ440695 LPZ440692 LPZ440632 LPZ440185 LPZ440107 LPZ440168 LPZ440046 LPZ440549
Root Cause
Software design
Action Taken
An URGENT MEDICAL DEVICE CORRECTION notification letter dated 9/18/25 was sent to customers. Actions to be taken by Customer/User For LOGIQ P9 and LOGIQ P10 series ultrasound systems with system software version R4.5.7, discontinue using the UGAP feature (please see Appendix for product identification). You may continue using other imaging features. After the correction has been implemented, please destroy the installation media for software R4.5.7 at your site. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records. Please complete and return the attached acknowledgement form to recall.74084@gehealthcare.com. GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to arrange for the correction. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.