RecallDepth

MEDLINE CLEARPRO TRACH T-PIECE CLOSED SUCTION CATHETER, flat peel pouch EA, 10 EA per Inner Carton (IC), 20 EA/2 Inner cartons per Case (CS). Labeled as follows: a. TRACH T-PIECE, 14 Fr, Item Number/REF: DYNCPTP14; b. TRACH T-PIECE, 14 Fr, Item Number/REF: DYNCPTP14T

Company
Medline Industries, Lp - Northfield
Recall Initiated
August 3, 2022
Posted
September 29, 2022
Recall Number
Z-1857-2022
Quantity
136,080 units
Firm Location
Northfield, IL

Reason for Recall

Potential for the suction catheter to come apart from the green connector near the thumb valve. The connector is coming apart from the assembled device when pulling the catheter out of the artificial airway in the patient suctioning process.

Distribution

Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and Puerto Rico. International distribution to Canada, Chile, Colombia, Panama, Qatar, United Arab Emirates

Lot / Code Info

a. UDI-DI (GTIN) case: (01)20888277703466; UDI-DI (GTIN) Each: (01)10888277703469, Lot Numbers 6922030011; b. UDI-DI (GTIN) case: (01)20888277703473; UDI-DI (GTIN) Each: (01)10888277703476, Lot Numbers 6921050011, 6921060011, 6921040012, 6921110011, 6921120011 6921120012, 6922010011, 6922010012, 6922010031, 6922020031, 6922030011.

Root Cause

Under Investigation by firm

Action Taken

On about 08/03/2022, Medline notified consignees via first class mail letter of the recall. The letter instructed consignees to immediately review inventory and destroy all affected product and complete the response form indicated at the link provided in the letter. Distributors were instructed to notify customers if further distributed, instruct those customers destroy any affected product, and include customer quantities on response form.