RecallDepth

Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 104-36-XX, MCS +5GXL Liner 5/15 DEG and MCS GXL Liner 5/15 DEG. Orthopedic hip implant component.

This recall is currently active, issued July 22, 2021. It was issued by Exactech, Inc..

Company
Exactech, Inc.
Recall Initiated
June 29, 2021
Posted
July 22, 2021
Recall Number
Z-2114-2021
Quantity
89,050 total
Firm Location
Gainesville, FL
Official Source
View on FDA website ↗

Reason for Recall

Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.

Distribution

Distribution US nationwide.

Lot / Code Info

All serial numbers. UDI-DI numbers: 10885862013682, 10885862013699, 10885862013705, 10885862013712, 10885862013729

Root Cause

Unknown/Undetermined by firm

Action Taken

The firm issued "URGENT MEDICAL DEVICE CORRECTION" letters dated 6/28/2021 via email on 6/29/2021 that were addressed to the attention of Exactech Agents, Representatives, and Distributors in possession of affected products.