RecallDepth

Exactech Equinoxe, Reverse Shoulder Humeral Liners, 322-38-00 (145-DEG PE 38MM HUM LINER +0), REF: 322-38-03 (145-DEG PE 38MM HUM LINER +2.5), REF:322-42-00 (145-DEG PE 42MM HUM LINER +0), REF: 322-42-03 (145-DEG PE 42MM HUM LINER +2.5)

Company
Exactech, Inc.
Recall Initiated
August 15, 2025
Posted
September 12, 2025
Recall Number
Z-2562-2025
Quantity
11,542 units
Firm Location
Gainesville, FL

Reason for Recall

Reverse Shoulder humeral liners have an articular surface position outside of the dimensional specification within the labeling.

Distribution

Worldwide - US Nationwide distribution in the states of CA, CO, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME, MN, NV, NY, OH, OK, PR, SC, TN, TX, VA, WI and the countries of Australia, Austria, Brazil, Canada, Colombia, France, United Kingdom, Guatemala, India, Italy, Japan, Korea, Germany, The Netherlands, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Dubai.

Lot / Code Info

All Serial Numbers, REF: 322-38-00/UDI:10885862593832, REF: 322-38-03/UDI:10885862593849, REF: 322-42-00/UDI: 10885862593917, REF:322-42-03/UDI:10885862593924

Root Cause

Process design

Action Taken

On August 15, 2025, Exactech, Inc. initiated a "Urgent Voluntary Medical Device Recall". Exactech asked consignees to take the following actions: 1. Healthcare providers who have treated patients using the Exactech Equinoxe Humeral Liners subject to this recall should continue to follow those patients pursuant to the healthcare provider s standard of care. 2. Do Not Use / Quarantine any of these products in your inventory. 3. Segregate and return unused units via prepaid packaging to Exactech. 4. Complete the Recall Confirmation Form confirming product return or on-hand inventory. 5. Transmission of this Recall Notice: This notice must be passed on to all who need to be aware within your organization and any locations the affected products has been transferred to. This recall has been reported to the U.S. FDA and will be reported to other regulatory authorities as required.