Medline Kits, trays, and packs labeled as follows: a) PORT ACCESS TRAY, REF DYNDA2863; b) PORT ACCESS TRAY, REF DYNDA2863H; c) ULTRASOUND GUIDED IV START KIT, REF DYNDV2518; d) ULTRASOUND GUIDED IV START KIT, REF DYNDV2518H; e) KIT ULTRASOUND PIV START, REF DYNDV2445; f) NICU INSERTION & DRSG CHG TRAY, REF EBSI1306A; g) PIV BSI KIT, REF BSIPIV1010; h) ULTRASOUND GUIDED IV START KIT, REF DYNDV2465

This recall is currently active, issued February 25, 2024. It was issued by Medline Industries, Lp - Northfield.

Company
Medline Industries, Lp - Northfield
Recall Initiated
November 15, 2023
Posted
February 25, 2024
Recall Number
Z-1113-2024
Quantity
3900 units
Firm Location
Northfield, IL
Official Source
View on FDA website ↗

Reason for Recall

Medline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.

Distribution

Distribution US nationwide including Puerto Rico, Canada, UAE, and Panama

Lot / Code Info

a) REF DYNDA2863, UDI/DI 40195327008025 (case), 10195327008024 (unit), Lot Numbers: 22ABQ171, 22ABT619, 22IBG373; b) REF DYNDA2863H, UDI/DI 40195327008025 (case), 10195327008024 (unit), Lot Numbers: 22ABQ171, 22ABT619, 22IBG373; c) REF DYNDV2518, UDI/DI 40195327089079 (case), 10195327089078 (unit), Lot Numbers: 22GBA169, 22HBU208; d) REF DYNDV2518H, UDI/DI 40195327089079 (case), 10195327089078 (unit), Lot Numbers: 22GBA169, 22HBU208; e) REF DYNDV2445, UDI/DI 20193489110576 (case), 10193489110579 (unit), Lot Numbers: 2022012480; f) REF EBSI1306A, UDI/DI 00653160352122 (case), 10653160352129 (unit), Lot Numbers: 2022042850; g) REF BSIPIV1010, UDI/DI 00653160349153 (case), 10653160349150 (unit), Lot Numbers: 2022020750, 2022040480, 2022082980; h) REF DYNDV2465, UDI/DI 20193489110163 (case), 10193489110166 (unit), Lot Numbers: 2022031480, 2022041880

Root Cause

Nonconforming Material/Component

Action Taken

MEDLINE issued an IMMEDIATE ACTION REQUIRED NURSE ASSIST PRODUCT RECALL notice to its consignees on 11/16/2023 via USPS first class mail and email. The notice explained the problem, requested log on to their portal to identify the products and check their inventory so the product could be quarantined. Once Medline receives the completed response form, over-labels will be shipped to the consignee. Medline Industries, LP instructed customers to over-label the convenience kits with a warning label indicating to remove the sodium chloride irrigation or Sterile water and replace from a different sterile supply. Distributors or consignees who further distributed the affected products were directed to notify their customers. For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.