Veran:SPiN Thoracic Navigation System Models: SYS-4230 SPiN Thoracic Navigation System - 230V 50Hz SYS-4000 SPiN Thoracic Navigation System SYS-3230 ig4 System, 230V, 50Hz / SPiN Thoracic Navigation System 0095 SYS-3000 SPiN Thoracic Navigation System SYS-2400 SPiN Drive System SYS-2230 ig4 System V2, 230V, 50Hz SYS-2000 ig4 Image Guides System SYS-1500 SPiN View" System SYS-1000 ig4 Image Guides System SYS-0220 ig4 System, 220V, 50Hz SYS-0185 SPiN Planning Laptop Workstation SYS-0128 Planning Workstation w/Monitor SYS-0002 ig4 System INS-7150 SPiN Vision Scope Adapter

This recall is currently active, issued September 29, 2023. It was issued by Olympus Corporation Of The Americas.

Company
Olympus Corporation Of The Americas
Recall Initiated
August 10, 2023
Posted
September 29, 2023
Recall Number
Z-2659-2023
Quantity
226 units
Firm Location
Center Valley, PA
Official Source
View on FDA website ↗

Reason for Recall

Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments) manufacturing to design requirements and insufficient evidence to demonstrate adequate electromagnetic compatibility (EMC). EMC is the ability of electrical equipment and systems to function acceptably without interference and to not interfere with other devices within its operating environment

Distribution

Worldwide distribution - US Nationwide and the countries of CN, DE, HK, IT, KR, SG, TH, TW.

Lot / Code Info

Model-UDI: SYS-4230 SPiN Thoracic Navigation System - 230V 50Hz 00815686020118; SYS-4000 SPiN Thoracic Navigation System 00815686020101; SYS-3230 ig4 System, 230V, 50Hz / SPiN Thoracic Navigation System 00815686020095; SYS-3000 SPiN Thoracic Navigation System 00815686020088; SYS-2400 SPiN Drive System N/A; SYS-2230 ig4 System V2, 230V, 50Hz 00815686020071; SYS-2000 ig4 Image Guides System 00815686020064; SYS-1500 SPiN View" System 00815686020057; SYS-1000 ig4 Image Guides System N/A; SYS-0220 ig4 System, 220V, 50Hz N/A; SYS-0185 SPiN Planning Laptop Workstation 00815686020033; SYS-0128 Planning Workstation w/Monitor N/A; SYS-0002 ig4 System N/A; INS-7150 SPiN Vision Scope Adapter N/A; All serial numbers and all lot numbers.

Root Cause

Under Investigation by firm

Action Taken

Olympus Corporation of the Americas ( Olympus ),[Veran Medical Technologies, a wholly owned subsidiary] issued Urgent Medical Device Recall Letter on 8/10/23 to: Endoscopy Department, Pulmonary Department, Risk Management for the VERAN ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments. Letter states reason for recall, health risk and action to take: Veran requests you to take the following actions: 1. Inspect your inventory and identify any Veran models. Please check all areas of the hospital to determine if any of these devices remain in inventory. 2. Call your Olympus customer service representative at 1-800-848-9024, option 2. Olympus will issue a Return Material Authorization to return any affected product at no charge. Your Olympus Customer Service representative will advise you on reimbursement for your Veran devices. 3. Olympus requests that you acknowledge receipt of this letter through the Olympus recall portal. a. Go to https://olympusamerica.com/recall b. Enter the file (recall) number: 0430 4. If you have further distributed this product, identify your customers, forward them this notification, and appropriately document your notification process. If you require additional information, please contact Tara Safi, Field Corrective Actions Administrator at Tara.Safi@Olympus.com.