Arjo Portable Scale Adaptor, Model Number 700.05725; accessory intended to be used in combination with compatible patient lifting equipment
- Company
- Arjohuntleigh Magog, Inc.
- Recall Initiated
- February 17, 2023
- Posted
- May 24, 2023
- Recall Number
- Z-1629-2023
- Quantity
- 3
- Firm Location
- Magog Canada
Reason for Recall
Some Arjo Portable Scale Adaptors have been assembled with straps that have a potential to detach.
Distribution
US: OH, NY
Lot / Code Info
UDI/DI 05055982760713, Serial Numbers: 300501841, 300501842, 300515661
Root Cause
Process control
Action Taken
Arjo issue an URGENT MEDICAL DEVICE RECALL notice to its consignees on 02/17/2023 via letter, UPS, 2nd Day Delivery. The notice explained the problem, the risk, and requested the consignee remove the units from use and destroy them. The firm will be replacing the units upon receipt of the completed response form. For questions or assistance in completing the Customer Response Form, please contact Arjo at 1 800-323-1245 or via email at ArjoCustomer.Response-US@arjo.com.