Modular Stem cylindrical, Tilastan, press-fit cementless, L= 80 mm, DIA = 11 mm, Product Code: 880-601/11
- Recall Initiated
- February 14, 2025
- Posted
- March 17, 2025
- Recall Number
- Z-1372-2025
- Quantity
- 37 units
Reason for Recall
A problem was discovered during the scanning process of the outer carton label because of an incorrect GTIN on the label, which is a GTIN for a different product.
Distribution
Worldwide - US Nationwide distribution in the states of AL, CA, CO, FL, GA, IN, KS, MO, NJ, NV, OH TX, VA, WI and the countries of Germany, Italy, Spain, Switzerland.
Lot / Code Info
Product Code: 880-601/11; UDI-DI: 04026575443925; Lot Numbers: 1943294, 1943296, 1943297, 2049174.
Root Cause
Labeling Change Control
Action Taken
On February 21, 2025 URGENT: MEDICAL DEVICE RECALL letters were sent to customers. Actions to be taken by the Customer/User: Please respond using the attached Distributor Reply form to indicate your understanding of this Device Recall and of the need to return the affected product. Pre-paid shipping labels will be provided once the Distributor Reply Form is returned. Please return the product as soon as possible to the following address: LinkBio Corporation 69 King St Dover, NJ 07801 Please contact LinkBio at 973-625-1333 or complaints@linkbio.com for any questions. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax.