MICRO-X Rover Mobile X-ray System, # MXU-RV19

This recall is currently active, issued December 12, 2024. It was issued by Micro-X Ltd. Unit 14 6 Mab Eastern Promenade Tonsley Australia.

Company
Micro-X Ltd. Unit 14 6 Mab Eastern Promenade Tonsley Australia
Recall Initiated
October 23, 2024
Posted
December 12, 2024
Recall Number
Z-0666-2025
Quantity
23
Official Source
View on FDA website ↗

Reason for Recall

During internal testing it was found that if the system is Ready to expose and the hand switch is then dropped into the stowage unit such that it lands on the prep/expose switch, the prep switch can be activated for a very short time such that the orange ring of death is displayed in the tube cooling area of the screen. The system then becomes unresponsive to kV/mAs changes, prep/expose presses. The only way to proceed is to suspend the patient, Exit Nexus DR and relogin and continue the exam.

Distribution

US Nationwide Distribution: TX, CA, OH, ID, WA, AZ

Lot / Code Info

UDI-DI: 09357123000013. Serial Numbers: 262, 274, 320, 329, 237, 269, 271, 290, 292, 307, 310, 311, 312, 321, 322, 323, 327, 330, 331, 332, 333, 234, 266.

Root Cause

Device Design

Action Taken

Starting on 10/23/24 Service Bulletin SB002-22 was sent to customers informing them that a field upgrade was released for the installation of an exposure switch holder. Customers were asked to complete and return an acknowledgement response form to support-us@micro-x.com For further information, contact your local service representative at support-us@micro-x.com, 1-206-249-8764.