A.L.P.S. mvX- MEDIAL TIBIA PLATE 16H LT-Medial Tibia Plate, 16-Hole, Left Intended for Fixation of fractures of the distal tibia Model/Catalog Number: 770715161

Company
Tyber Medical
Recall Initiated
January 21, 2025
Posted
February 21, 2025
Recall Number
Z-1225-2025
Quantity
100 units
Firm Location
Bethlehem, PA

Reason for Recall

Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices

Distribution

Nationwide

Lot / Code Info

UDI-DI 00196449016037; Lot Numbers 24061DF01, 24065DF02, 24070DF01

Root Cause

Process control

Action Taken

Tyber Medical issued Urgent Medical Device Recall letter on 1/24/25 via email to the Distributor. Letter states reason for recall, health risk and action to take: 1. Acknowledge the receipt of this communication via email correspondence to feedback@tybermed.com. 2. Further distribute this communication to all applicable parties and/or customers that possess the impacted inventory. 3. Identify via laser-marked lot number and quarantine the impacted plates under Zimmer Biomet s control that have not been distributed for sale. An RMA will be provided for the return of goods to Tyber Medical for further inspection. 4. Identify via laser-marked lot number and quarantine the impacted plates that have been distributed. An RMA will be provided for the return of goods to Tyber Medical for further inspection. 5. Complete the customer response form located in Appendix 2 and return it to Tyber Medical via email correspondence to feedback@tybermed.com.

More recalls by Tyber Medical

View all recalls by this company →