RecallDepth

Brand Name: VOLT" Wrist Treatment System Product Name: 2.4/2.7 2Column Dist Radius Plate 6Hd/10Shaft/Left/Std/139mm Model/Catalog Number: 02.426.681S Product Description: 2 Column Distal Radius Plate - 6 Head, 10 Shaft, Standard Left - 139mm

Company
Tyber Medical
Recall Initiated
December 19, 2025
Posted
January 20, 2026
Recall Number
Z-1131-2026
Quantity
100 units
Firm Location
Bethlehem, PA

Reason for Recall

The supplier manufactured anatomical left plates with an incorrect thread orientation.

Distribution

US Nationwide distribution in the state of PA.

Lot / Code Info

Model No: 02.426.681S; UDI-DI: 00196449020508; Lot Number: 291294;

Root Cause

Process control

Action Taken

On December 19, 2025, an URGENT MEDICAL DEVICE REMOVAL NOTIFICATION email was sent to their only customer. Actions to be taken: 1. Acknowledge the receipt of this communication via email correspondence to WeCare@exalta.com 2. Further distribute this communication through notifying the accounts that have received the identified devices. 3. Identify via product label and quarantine the impacted devices under DePuy Synthes control all devices that have not been distributed for sale. An RMA will be provided for the return of goods to Tyber Medical. 4. Identify via product label and quarantine the impacted devices that have been distributed. An RMA will be provided for the return of goods to Tyber Medical. 5. Complete the customer response form following this notification, or provide written notification, and return it to Tyber Medical via email correspondence to WeCare@exalta.com. A sub-notification letter was also provided to the distributor to provide to customers with the following actions: 1. Identify via product label and quarantine the impacted devices in your possession. 2. Complete the subsequent Acknowledgement and Receipt From and return it to DePuy Synthes via email correspondence, with Tyber Medical copied at WeCare@exalta.com 3. Return the impacted product to DePuy Synthes for replacement.

More recalls by Tyber Medical

View all recalls by this company →