Duravent Silicone Ventilation Tube. Model Number: 240075.

This recall is currently active, issued December 19, 2024. It was issued by Olympus Corporation Of The Americas.

Company
Olympus Corporation Of The Americas
Recall Initiated
November 19, 2024
Posted
December 19, 2024
Recall Number
Z-0707-2025
Quantity
984 units
Firm Location
Center Valley, PA
Official Source
View on FDA website ↗

Reason for Recall

Potential that Duravent Silicone Ventilation Tubes may contain incorrect product, specifically, 1.32 mm Tiny T Tubes (model: 70240076) instead of the intended Duravent Silicone Ventilation Tubes.

Distribution

AL, AZ, CA, IL, KY, MD, MN, MO, NC, ND, NY, OK, PA, SC, TN, TX, WA, WI. International distribution to Australia, Canada, Germany.

Lot / Code Info

Model Number: 240075. UDI: 00521925031883 (17)331015 (10)SM405666, Lot Number: SM405666. UDI: 00521925031883 (17)331015 (10)JB362646, Lot Number: JB362646.

Root Cause

Process control

Action Taken

Olympus notified consignees on 11/19/2024 via Courier, 2-day delivery. Consignees were instructed to examine inventory for affected devices, quarantine any affected devices on hand, arrange for return of those affected units and acknowledge receipt of the customer letter through the web portal.