Medline brand medical procedure convenience kits, labeled as: a) LITHOTOMY PACK, REF DYNJ24362C; b) MAJOR GYNE LAPAROSCOPY SHSC, REF DYNJ46774B; c) GYN ANCILARY PACK, REF DYNJ84509; d) GYN LITHOTOMY MINOR, REF DYNJ907707
This recall is currently active, issued October 23, 2024. It was issued by Medline Industries, Lp - Northfield.
- Recall Initiated
- September 4, 2024
- Posted
- October 23, 2024
- Recall Number
- Z-0175-2025
- Quantity
- 197 units
- Firm Location
- Northfield, IL
- Official Source
- View on FDA website ↗
Reason for Recall
It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Panama.
Lot / Code Info
a) REF DYNJ24362C, UDI/DI 40889942830054 (each), 10889942830053 (case), Lot Numbers: 24ABG207; b) REF DYNJ46774B, UDI/DI 40889942582748 (each), 10889942582747 (case), Lot Numbers: 23LBN531; c) REF DYNJ84509, UDI/DI 40195327372881 (each), 10195327372880 (case), Lot Numbers: 23LBL712; d) REF DYNJ907707, UDI/DI 40193489848978 (each), 10193489848977 (case), Lot Numbers: 24ABR224
Root Cause
Under Investigation by firm
Action Taken
Medline issued a MEDICAL DEVICE RECALL notice to its consignees on 09/05/2024 via email and USPS. The notice explained the issue, hazard, and requested the destruction of the affected product. Distributors were directed to notify their consignees.