Medline convenience kits containing semi-rigid suction liners labeled as follows: KIT NEURO SHUNT ST LOUIS CHILD, Pack Number DYNJ907829B
This recall is currently active, issued September 11, 2024. It was issued by Medline Industries, Lp - Northfield.
- Recall Initiated
- June 27, 2024
- Posted
- September 11, 2024
- Recall Number
- Z-3117-2024
- Quantity
- 134 units
- Firm Location
- Northfield, IL
- Official Source
- View on FDA website ↗
Reason for Recall
Due to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Panama.
Lot / Code Info
DYNJ907829B: UDI/DI case 40195327077991, UDI/DI each 10195327077990, Lot Numbers: 23LMA774, 23LME672, 24AMA991, 24CMB554, 24CMB648
Root Cause
Process control
Action Taken
Medline Industries issued a MEDICAL DEVICE RECALL notice to its consignees on 06/27/2024 via US mail and email. The notice explained the problem with the device, potential risk to the patient, and requested the consignee immediately check their stock for the affected item number and the affected lot numbers and destroy any affected inventory. Distributors were directed to notify their customers of the recall and request the affected products be destroyed. For questions: contact the Recall Department at 866-359-1704 or recalls@medline.com.