Toshiba Medical Kalare Fluoroscopic X-Ray System (DREX-KL80) Kalare is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic examinations, including general R&F and pediatric examinations. It is intended for use by qualified/trained doctors or technologists on both adult and pediatric subjects taking radiographic and fluoroscopic exposures of the whole body, skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing or lying in prone or supine positions.

This recall has been terminated (originally issued September 12, 2017).

Company
Toshiba American Medical Systems Inc
Recall Initiated
September 12, 2017
Terminated
November 18, 2020
Recall Number
Z-0012-2018
Quantity
US - 96
Firm Location
Tustin, CA
Official Source
View on FDA website ↗

Reason for Recall

During an examination images were displayed on the live monitor,but the images were not displayed on the system monitor nor were they saved to the hard disk.

Distribution

USA (nationwide) Distribution

Lot / Code Info

Model: HDR-08A

Root Cause

Software design

Action Taken

Toshiba planned action to bring defect into compliance: 1. You will issue a Customer Notification Letter to make customers aware of the hazards, what measures will be taken to remedy the defect at no cost to the user, and how to safely use the equipment before corrections can be made. 2. You will install modified firmware on the systems to correct the defect and prevent recurrence. 3. You will provide progress reports on the effectiveness of the plan. CDRH approves the CAP subject to the following conditions: 1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator are to be included in the notification. Thank you for your urgent attention to this matter. If you have any questions regarding this letter please feel free to call (800) 421-1968, or contact your local Toshiba Representative at (800) 521-1968. For further questions please call (714) 730-5000.