RecallDepth

GrebSet Micro-Introducer Kit, model 7941, Sterilized with Ethylene Oxide. Product Usage: The GrebSet micro-introducer kit is intended to facilitate the percutaneous placement of guidewires in the vascular system and for delivery of contrast media to vascular sites.

This recall has been terminated (originally issued January 12, 2017).

Company
Vascular Solutions, Inc.
Recall Initiated
January 12, 2017
Terminated
April 18, 2018
Recall Number
Z-1250-2017
Quantity
633
Firm Location
Maple Grove, MN
Official Source
View on FDA website ↗

Reason for Recall

Four Lots of GrebSet Micro-introducer Kits, model 7941, contain guidewires that have a shelf life shorter than the kit expiration date.

Distribution

US Nationwide Distribution in the states of NH, ME, VT, NY, LA, MS, AL CA, IA, WA, OR, MD, DE, PA, NJ, ID, TX, OK, MO, WI, GA, FL, AR, OH, AZ.

Lot / Code Info

Lot Numbers: 581212, 584392, 594115, 597202

Root Cause

Error in labeling

Action Taken

Consignees were sent on 1/12/2017 a Vascular Solutions "Urgent Medical Device Recall" letter dated January 11th, 2017. The letter identified the problem and product involved in the recall. The letter provided "Your Immediate Action Is Required". This included to remove the product from inventory and to complete and return the VSI's Account Inventory Form to VSIs Customer Service Dept.

More recalls by Vascular Solutions, Inc.

View all recalls by this company →