Langston Dual Lumen Catheter

This recall has been terminated (originally issued October 20, 2020).

Company
Vascular Solutions, Inc.
Recall Initiated
August 31, 2020
Posted
October 20, 2020
Terminated
June 18, 2024
Recall Number
Z-0110-2021
Quantity
1,680 units
Firm Location
Maple Grove, MN
Official Source
View on FDA website ↗

Reason for Recall

The inner lumen of the Langston catheter may separate from the hub during pressure injection.

Distribution

Domestic Distribution: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL ,GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ,NM ,NV ,NY ,OH ,OK,OR, PA ,PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY. International Distribution: Austria, Australia, Belgium, Bulgaria, Canada, Cyprus, Czech Republic, France, Germany, Greece, Hong Kong, Ireland, Israel, Italy, Mexico, Netherlands, Oman, Poland, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, United Arab Emirates and United Kingdom.

Lot / Code Info

Model Number: 5515, UDI 10841156100506 Lot Numbers: 635806, 640206, 644140, 648701, 654143, 658252, 659635, 661470, 665301, 670376, 654130, 672728, 656020, 638111, 677058.

Root Cause

Process control

Action Taken

The firm notified consignees that received affected product on August 26, 2020 via UPS Mail (US), email or mail (OUS). The firms immediate correction is to recall the affected devices for destruction. The notice includes instructions on subrecall for any consignees that further distributed affected product.