smith&nephew ANTHOLOGY POROUS HIGH OFFSET STEM, 12/14 TAPER, 7H SIZE, 32 MM x 112 MM, REF 71356107 Hip prosthesis.

This recall has been terminated (originally issued May 9, 2018).

Company
Smith & Nephew Inc
Recall Initiated
May 2, 2018
Posted
May 9, 2018
Terminated
September 1, 2020
Recall Number
Z-1921-2018
Quantity
13 units
Firm Location
Memphis, TN
Official Source
View on FDA website ↗

Reason for Recall

Three lots of Anthology HO Porous size 7 stems were manufactured with an undersized taper length. The device taper length was incorrectly manufactured as 0.393mm opposed to 0.442 mm.

Distribution

US nationwide distribution.

Lot / Code Info

Lot Numbers: 17LM22019, 17LM10934, 17LM05900

Root Cause

Process control

Action Taken

The firm initiated the recall by letter on 05/02/2018. Consignees are asked to inspect inventory, locate and quarantine affected, unused devices. Affected product will be returned to Smith & Nephew. All consignees are asked to complete the Response Form, even if they do not have product to return. Distributors were directed to notify their customers of the field action and ensure that these actions are carried out. For further questions, please call (978) 749-1440