3M Ranger(TM) Pressure Infusor, Model 145, SKU 70200792979 The 3M Ranger Pressure Infusor is a hardware device intended to provide pressure to I.V. solution bags when rapid infusion of liquids is required.

Company
3M Company Health Care Business 3M Center
Recall Initiated
May 22, 2017
Terminated
June 14, 2018
Recall Number
Z-2632-2017
Quantity
1 unit
Firm Location
Saint Paul, MN

Reason for Recall

It was discovered that one Ranger Pressure lnfusor Model 145, SN: 100873, produced a left chamber pressure that was slightly below the manufacturing specification.

Distribution

U.S. Distribution: TX only

Lot / Code Info

Serial Number: 100873

Root Cause

Process control

Action Taken

The firm, 3M, sent a letter dated 5/22/2017, to its consignee/customer. The letter described the product, problem and actions to be taken. The consignee was informed that there is no safety issue; however, the firm will be replacing unit. In addition, a representative from 3M will be visiting the facility and exchanging the product for a new one. Please feel free to call (713) 292-4430 or email baprestera@mmm.com if you have any questions.