RecallDepth

cobas p 612 pre-analytical system; Pre-analytical sample handling that includes de-capping, Aliquotting and sorting of samples for analysis.

This recall has been terminated (originally issued January 24, 2017).

Company
Roche Diagnostics Corporation
Recall Initiated
December 27, 2016
Posted
January 24, 2017
Terminated
August 4, 2017
Recall Number
Z-1090-2017
Quantity
5
Firm Location
Indianapolis, IN
Official Source
View on FDA website ↗

Distribution

TX and WA

Lot / Code Info

Serial Number 63004316 63004416 63004516 63002816 63002916

Root Cause

Software design

Action Taken

On 12/27/2016, URGENT MEDICAL DEVICE CORRECTION notifications were sent to the affected Consignees via courier fax or email. The recall notifications included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. The cobas p 612 pre-analytical system (63x) software is being revised to correct the aspiration process. We anticipate this new version (2.0.4) to be available during Quarter 1 2017. For now, discontinue using disposable non-filter tips for Aliquotting on your cobas p 612 pre-analytical system (63x). You will be contacted by a Roche Field Engineering Specialist (FES) about modifying your system to allow for the use of disposable filter tips. Filter tips will be provided at the time of service. Actions Required " Immediately discontinue using disposable non-filter tips for Aliquotting on your cobas p 612 pre-analytical system (63x). " Complete the enclosed faxback form and fax it to number 1-877-906-8982 or email it to Indianapolis.cd_tsc@roche.com. " Provide copies of this UMDC to other clinicians who may need to be aware of this software issue. " If your facility has distributed the affected product to another site, please ensure this UMDC is provided to that site. " File this UMDC for future reference. Please contact the Roche Support Network Customer Support Center at 1-800-428-2336, 24 hours a day/7 days a week, if you have questions about the information contained in this UMDC.

More recalls by Roche Diagnostics Corporation