Medtronic Bio-Console 560 Extracorporeal Blood Pumping Console, Model Numbers: 560BC, 560BC1, 560BCS, 560BCS1, R560BCS1. Used in cardiopulmonary bypass (CPB) procedures.
- Company
- Medtronic Perfusion Systems
- Recall Initiated
- March 3, 2021
- Recall Number
- Z-1313-2021
- Quantity
- 387 devices (93 US)
- Firm Location
- Brooklyn Park, MN
Reason for Recall
An electrical component within the system controller module of certain Bio-Console 560 devices may exhibit electrical failure causing the device to stop pumping. Issues observed include the pump stopping, the user interface going blank, and smoking and producing a burning odor during use.
Distribution
Worldwide distribution. US Nationwide including Puerto Rico, Brazil, Brunei Darussalam, Canada, China, Colombia, Croatia, Ecuador, Guatemala, India, Israel, Italy, Japan, Kazakhstan, Korea, Mexico, Morocco, New Zealand, Russian Federation, Spain, Taiwan, Thailand, United Kingdom, and Viet Nam
Lot / Code Info
560BC: GTIN 00885074539386, Serial Numbers: 560B100094, 560B100212, 560B100244, 560B100257, 560B100274, 560B100283, 560B100418, 560B100439, 560B100482 560BC1: GTIN 00613994450326, Serial Numbers: 560B100901, 560B101368 560BCS: GTIN 00613994258236, Serial Numbers: 560B100369, 560B100376, 560B100472, 560B100495, 560B100496 560BCS1: GTIN: 00613994450463, Serial Numbers: 560B100191, 560B100690, 560B100783, 560B100806, 560B100820, 560B100871, 560B100896, 560B101062, 560B101187, 560B101318, 560B101344, 560B101470, 560B101499, 560B101504, 560B101519, 560B101593, 560B101526 GTIN 00613994526052, Serial Numbers: 560B102610, 560B102725 GTIN 00643169201033, Serial Numbers: 560B101838 GTIN 00643169201651, Serial Numbers: 560B101834 GTIN 00643169315822, Serial Numbers: 560B100127, 560B101694, 560B101700, 560B101805, 560B101976, 560B101996, 560B102080, 560B102118, 560B102185, 560B102194, 560B102204, 560B102210, 560B102215, 560B102265, 560B102336, 560B102426, 560B102493, 560B102518, 560B102547, 560B102559, 560B102598, 560B102761, 560B102762, 560B102763, 560B102764, 560B102765, 560B102766 GTIN 00763000135447, Serial Numbers: 560B102711, 560B102717, 560B102718, 560B102720, 560B102721, 560B102722, 560B102723, 560B102724, 560B102726, 560B102727, 560B102730, 560B102731, 560B102732, 560B102733, 560B102734, 560B102735, 560B102736, 560B102737, 560B102738, 560B102739, 560B102740, 560B102741, 560B102742, 560B102743, 560B102745, 560B102746, 560B102747, 560B102748, 560B102750, 560B102751, 560B102752, 560B102754, 560B102755, 560B102756, 560B102757, 560B102758, 560B102760, 560B102767, 560B102768, 560B102769, 560B102770, 560B102771, 560B102772, 560B102773, 560B102774, 560B102775, 560B102776 GTIN 00763000159412, Serial Numbers: 560B102930, 560B102931, 560B102932, 560B102933, 560B102934 GTIN 00763000236526, Serial Numbers: 560B102778, 560B102779, 560B102780, 560B102781, 560B102782, 560B102783, 560B102784, 560B102785, 560B102786, 560B102787, 560B102788, 560B102789, 560B102790, 560B102792, 560B102793, 560B102794, 560B102795, 560B102796, 560B102797, 560B102798, 560B102799, 560B102800, 560B102801, 560B102802, 560B102803, 560B102804, 560B102805, 560B102806, 560B102807, 560B102809, 560B102810, 560B102811, 560B102812, 560B102813, 560B102815, 560B102817, 560B102818, 560B102819, 560B102820, 560B102821, 560B102822, 560B102823, 560B102824, 560B102825, 560B102826, 560B102827, 560B102828, 560B102829, 560B102830, 560B102831, 560B102832, 560B102833, 560B102834, 560B102835, 560B102836, 560B102837, 560B102838, 560B102839, 560B102840, 560B102841, 560B102842, 560B102843, 560B102844, 560B102845, 560B102846, 560B102847, 560B102848, 560B102849, 560B102850, 560B102851, 560B102852, 560B102853, 560B102854, 560B102855, 560B102856, 560B102857, 560B102858, 560B102859, 560B102860, 560B102861, 560B102862, 560B102863, 560B102864, 560B102865, 560B102866, 560B102867, 560B102868, 560B102869, 560B102870, 560B102871, 560B102872, 560B102873, 560B102874, 560B102875, 560B102876, 560B102877, 560B102878, 560B102879, 560B102880, 560B102882, 560B102883, 560B102884, 560B102885, 560B102886, 560B102887, 560B102888, 560B102889, 560B102890, 560B102891, 560B102892, 560B102893, 560B102894, 560B102895, 560B102896, 560B102897, 560B102898, 560B102899, 560B102900, 560B102901, 560B102902, 560B102903, 560B102907, 560B102908, 560B102909, 560B102910, 560B102912, 560B102913, 560B102914, 560B102915, 560B102916, 560B102917, 560B102918, 560B102919, 560B102920, 560B102921, 560B102922, 560B102923, 560B102924, 560B102925, 560B102926, 560B102927, 560B102928, 560B102929, 560B102946, 560B102947, 560B102948, 560B102949, 560B102950, 560B102951, 560B102952, 560B102953, 560B102954, 560B102955, 560B102956, 560B102957, 560B102958, 560B102959, 560B102960, 560B102961, 560B102962, 560B102963, 560B102964, 560B102965, 560B102966, 560B102967, 560B102968, 560B102969, 560B102970, 560B102971, 560B102972, 560B102973, 560B102974, 560B102975, 560B102976, 560B102977, 560B102978, 560B102979, 560B102980, 560B102981, 560B102982, 560B102983, 560B102984, 560B102985, 560B102986, 560B102987, 560B102988, 560B102989, 560B102990, 560B102991, 560B102992, 560B102993, 560B102994, 560B102995, 560B102996, 560B102997, 560B102998, 560B102999, 560B103000, 560B103001, 560B103002, 560B103003, 560B103004, 560B103005, 560B103006, 560B103007, 560B103008, 560B103009, 560B103010, 560B103011, 560B103012, 560B103013, 560B103014, 560B103015, 560B103016, 560B103017, 560B103018, 560B103019, 560B103020, 560B103021, 560B103022, 560B103023, 560B103024, 560B103025, 560B103026, 560B103027, 560B103028, 560B103029, 560B103031, 560B103032, 560B103033, 560B103034, 560B103035, 560B103036, 560B103037, 560B103038, 560B103039, 560B103040, 560B103041, 560B103042, 560B103043, 560B103044, 560B103045, 560B103046, 560B103047, 560B103048, 560B103049, 560B103050, 560B103051, 560B103052, 560B103053, 560B103054, 560B103055, 560B103057, 560B103058, 560B103059, 560B103060, 560B103061, 560B103062, 560B103069, 560B103070, 560B103071, 560B103072, 560B103073, 560B103074, 560B103075, 560B103077, 560B103078, 560B103079 R560BCS1, GTIN 00643169315822, Serial Numbers: 560B102009, 560B102355, 560B102367
Root Cause
Component change control
Action Taken
In the United States (US), beginning 03-Mar-2021, a written Urgent Medical Device Correction letter was mailed via UPS 2-day delivery to consignees who have received affected product according to Medtronic records. Consignees will be asked to complete and return a confirmation certificate, confirming they have received the notification.