Octopus Nuvo Tissue Stabilizer, disposable tissue stabilization system, Product Number TSMICS1
- Company
- Medtronic Perfusion Systems
- Recall Initiated
- January 13, 2026
- Posted
- February 26, 2026
- Recall Number
- Z-1473-2026
- Quantity
- 2 units
- Firm Location
- Brooklyn Park, MN
Reason for Recall
During the manufacturing, Medtronic personnel identified an assembly issue where the tubing was incorrectly positioned. The short tube was attached to the bottom of the canister, and the long tube was attached to the top, contrary to the intended design.
Distribution
US Nationwide distribution in the states of FL and IN.
Lot / Code Info
UDI-DI: 00763000543693; Lot Number: 0232837110
Root Cause
Under Investigation by firm
Action Taken
An URGENT: MEDICAL DEVICE RECALL notification letter dated January 2026 was hand delivered to customers. Actions: Medtronic requests that you take the following actions: " Review your inventory for listed lot number. " If you have inventory from the listed lot number immediately quarantine and return it to Medtronic by contacting Customer Service at 1-800-854-3570, Option 1 then Option 4, and referencing this communication to initiate return and exchange or credit of this unit. Your Medtronic sales representative can assist you in the return of unit as necessary. " Complete the enclosed Customer Confirmation Form and email to RS.CFQFCA@medtronic.com. This form must be returned even if you do not have the affected product in your possession. " Please maintain a copy of this communication in your records. If you have any questions regarding this communication, please contact your Medtronic Field Representative.