Fellows "Mighty 8" Power strip (Part Number 04175-0008) and TrippLite Power strip (PS-415-HGULTRA - Part Number 13-00106-00) are two components sold as part of the Merge Hemo System.

This recall has been terminated (originally issued April 6, 2017).

Company
Merge Healthcare, Inc.
Recall Initiated
April 6, 2017
Terminated
July 10, 2019
Recall Number
Z-0350-2018
Quantity
1,150 power strips
Firm Location
Hartland, WI
Official Source
View on FDA website ↗

Reason for Recall

Merge is conducting a hardware recall (replacement) of Fellows "Mighty 8" Power Strip and TrippLite Power Strip due to overheat of input resistor causing the power strip to smoke.

Distribution

Nationwide.

Lot / Code Info

Part Number 04175-0008 and Part Number 13-00106-00

Root Cause

Component design/selection

Action Taken

Merge began notify customers via an Urgent Medical Device Recall letter on 4/6/2017 via mail. The letter identified affected product, stated the issue, and informed customers that Merge would replace affected product. A response form is requested to be returned to Merge. Questions can be sent via email to recall@merge.com