Radial Osteotome Blade, REF 71369314. Used for cutting or preparing bone.

This recall has been terminated (originally issued October 23, 2017).

Company
Smith & Nephew Inc
Recall Initiated
October 23, 2017
Terminated
August 24, 2018
Recall Number
Z-0406-2018
Quantity
9 units
Firm Location
Memphis, TN
Official Source
View on FDA website ↗

Reason for Recall

The expiration date provided on the label is incorrect.

Distribution

Worldwide Distribution -- US, to the states of TN and NJ; and, Australia.

Lot / Code Info

Lot Numbers: 14HSY0006R, 15MSY0005R, 15KSY0023R, 09FSY0003R, and 12HSY0040R.

Root Cause

Process change control

Action Taken

The firm initiated their recall by an Urgent Medical Device Recall Notice on 10/23/2017. The letter identified the affected device and the reason for the recall, as well as the potential risk. The firm instructed users to please inspect their inventory and locate any unused devices from the list of product and batch numbers provided in the Notice. Affected devices must be quarantined immediately. Distributors, must notify their customers of the field action and ensure that the actions are carried out. The attached form should be completed and returned. Questions or concerns can be directed to FieldActions@smith-nephew.com.