4205 Journey Rolling Walker Product Usage: used to assist people in walking

Company
Nova Ortho-Med Inc
Recall Initiated
October 1, 2016
Posted
October 24, 2016
Terminated
February 6, 2017
Recall Number
Z-0304-2017
Quantity
1,380 units
Firm Location
Carson, CA

Reason for Recall

Nova Medical is recalling the Journey Rolling Walker because the screws used to hold the black side support pieces to the frame can become loose over time. This can result in the locking nut coming off and the screw falling out.

Distribution

US Nationwide Distribution in the states of: CA, IN, HI, CT, UT, WA, FL, VA, WI, OH, MN, KS, MD, IL, MO, WY, MT, ME, PA, TN, MI, OR, NY, NJ, ID, NC, IA, OK, AR, VT, SC, TX

Lot / Code Info

Serial No. YA 4205RDFG004 TO YA4205RDFG0098 YZ4205BKFC0002 TO YZ4205RDFE0100

Root Cause

Component design/selection

Action Taken

A customer notification letter dated 10/1/16 was sent to customers to inform them that the screws used to hold the black side support pieces to the frame can become loose over time. This can result in the locking nut coming off and the screw falling out. This allows the frame support piece to detach from the walker and causes instability resulting in possible injury. The letter informs the customers of the root cause and the solution. Customers are instructed to record the serial #s of the product that have been corrected and email them to NOVA Customer Service (patricias@novajoy.com). Customers can also 1-800-557-6682 x131.