Star 8 Rolling Walker/Rollator, Items: 4258/ BK(Black)/ BL(Blue)/ DB(Sky Blue)/ PL(Purple)/ RD(Red)

Company
Nova Ortho-Med Inc
Recall Initiated
October 7, 2019
Recall Number
Z-0654-2021
Quantity
10898
Firm Location
Carson, CA

Reason for Recall

Walkers within the identified serial range have the potential for the front fork to unscrew itself from the leg. A partially unscrewed fork creates unevenness in the wheelbase, which could cause instability, that if not detected by the user, may result in the front fork falling off the walker.

Distribution

U.S.: AK, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, MI, MN, MO, MT, NC, NE, NJ, NM, NV, OH, OK, OR, PA, SC, SD, TN, TX, UT, WA, WI, WY O.U.S.: Taiwan

Lot / Code Info

Serial # range: HF-4258BKHA0001 to HF4258RDHK0352 (Purchased before April 4th,2019)

Root Cause

Device Design

Action Taken

On 10/07/2019, Product Correction and Product Safety Recall notices were emailed to customers. Customers were asked to replace the front leg assemblies as follows: a. Depress the push buttons on the original legs and remove them from the frame. b. Insert the improved replacement assemblies issued by NOVA into the unit and discard the original legs. In addition, customers were asked to do the following: 1. Check your inventory for affected walkers and fix. 2. Contact all customers who purchased affected walkers notify them of the potential issue. Schedule a time that they can come in for repair or if the customer prefers the recalling firm can send the wheel assemblies directly to their residence. Have them inspect their walker immediately to identify if either of the front forks have unscrewed from the leg. If so, please instruct them to not use the walker until the leg assemblies can be replaced with the new parts. 3. Call the recalling firm to request the wheel assembly set to be sent to the required location. 4. Record the serial #'s of the product that have been corrected and email them to Customer Service (info@novajoy.com) so that we can have for our records. If you prefer to call, then dial 1-800-557-6682 x8129. Email: info@novajoy.com

More recalls by Nova Ortho-Med Inc

View all recalls by this company →