Beacon EUS Access System Preloaded Access 90degree Needle sterile, single-use, endoscopic ultrasound device consisting of a sharp stylet and an 18.5-gauge, echo-enhanced, tip-shaped, nitinol cannula intended to pierce tissue and then act as an access channel to facilitate guidewire placement to locations of the gastrointestinal tract.
- Company
- Covidien Llc
- Recall Initiated
- June 26, 2020
- Posted
- August 5, 2020
- Recall Number
- Z-2768-2020
- Quantity
- 375 units
- Firm Location
- Mansfield, MA
Reason for Recall
This voluntary recall is being conducted following three customer reports of the device cannula detaching during a procedure. In one case, the detached cannula was not retrieved during the procedure. Potential risks associated with a detached cannula include delay in treatment, biliary leak, pain and retained device fragments.
Distribution
world wide distribution
Lot / Code Info
Model # DSA-090-01 UDI: 10884521722583 Lot #: F2510370X, F2510900X, F2512340X, F2512378X, F2512492X, F2512620X, F2513452X
Root Cause
Under Investigation by firm
Action Taken
Medtronic sent a letter notification to customers dated July 06, 2020 recalling effected lots of product. This letter instructed customers to discontinue use of the product and quarantine effected inventory. Customers were also instructed to return effected product and complete recall response forms accordingly.