JOURNEY II UNI RESECTION PREP TRIAL, RIGHT MEDIAL/ LEFT LATERAL - Product Usage: The JOURNEY II Uni Resection Prep Trial is a reusable surgical instrument used during range of motion trialing to assess joint tension and component alignment during surgery. The provisional moveable spikes (spike rail) aids in provisional fixation of the trial. SIZE 1, CATALOG NUMBER: REF 74035831 SIZE 2, CATALOG NUMBER: REF 74035832 SIZE 3, CATALOG NUMBER: REF 74035833 SIZE 4, CATALOG NUMBER: REF 74035834 SIZE 5, CATALOG NUMBER: REF 74035835 SIZE 6, CATALOG NUMBER: REF 74035836 SIZE 7, CATALOG NUMBER: REF 74035837 SIZE 8, CATALOG NUMBER: REF 74035838 SIZE 9, CATALOG NUMBER: REF 74035839 SIZE 10, CATALOG NUMBER: REF 74035840
This recall has been terminated (originally issued January 28, 2020).
- Company
- Smith & Nephew Inc
- Recall Initiated
- January 28, 2020
- Terminated
- December 11, 2023
- Recall Number
- Z-2133-2020
- Quantity
- 195 units
- Firm Location
- Memphis, TN
- Official Source
- View on FDA website ↗
Reason for Recall
The spike rail was manufactured with an oversized thru-hole which could potentially cause the dowel pin to fall out and the device to disassemble.
Distribution
US Nationwide distribution.
Lot / Code Info
SIZE 1, CATALOG NUMBER: REF 74035831; LOT/BATCH CODES: 19FM01064R, 19HM18082, 19HM21871, 19JM18534, 19JM21086, 19KM07281, 19KM15577 SIZE 2, CATALOG NUMBER: REF 74035832; LOT/BATCH CODES: 19EM21587R, 19FM11181R, 19HM21899, 19HM21902, 19HM21903, 19JM21089 SIZE 3, CATALOG NUMBER: REF 74035833; LOT/BATCH CODES: 19GM21768, 19HM18085, 19HM18085A, 19JM09477, 19JM18537, 19JM18540, 19JM18541, 19JM18542, 19JM18542A SIZE 4, CATALOG NUMBER: REF 74035834; LOT/BATCH CODES: 19FM11183R, 19JM07474, 19JM11137, 19JM21090, 19KM21513 SIZE 5, CATALOG NUMBER: REF 74035835; LOT/BATCH CODES: 19FM01065R, 19HM18086, 19HM18086A, 19HM18088, 19JM18491, 19JM21091, 19KM15580 SIZE 6, CATALOG NUMBER: REF 74035836; LOT/BATCH CODES: 19FM11184R, 19GM21778, 19JM07477, 19JM21093 SIZE 7, CATALOG NUMBER: REF 74035837; LOT/BATCH CODES: 19FM11185A, 19FM11185R, 19HM19006, 19JM21096, 19KM21504, 19KM21506, 19KM21509 SIZE 8, CATALOG NUMBER: REF 74035838; LOT/BATCH CODES: 19FM11191R, 19FM11191S, 19HM18091, 19HM21040, 19JM18543, 19JM21097, 19JM23714, 19KM15583 SIZE 9, CATALOG NUMBER: REF 74035839; LOT/BATCH CODES: 19FM05578R, 19HM18096, 19HM19658, 19KM05613, 19KM15584, 19KM15586 SIZE 10, CATALOG NUMBER: REF 74035840; LOT/BATCH CODES: 19HM18107, 19HM19004, 19JM23666, 19JM23674, 19JM23678, 19JM23678A, 19JM23680, 19KM12708
Root Cause
Process design
Action Taken
The firm initiated the recall by email on 01/28/2020. The initial consignee notified the district sales representatives on 01/28/2020 by email. The district sales representatives checked their inventory and the inventory of the hospital they service.