Elecsys BRAHMS PCT Immunoassay for the in vitro quantitative determination of PCT (procalcitonin) in human serum and plasma (K2 and K3 EDTA, Li-Heparin).

This recall has been terminated (originally issued December 22, 2017).

Company
Roche Diagnostics Corporation
Recall Initiated
December 22, 2017
Terminated
April 5, 2019
Recall Number
Z-1311-2018
Quantity
47
Firm Location
Indianapolis, IN
Official Source
View on FDA website ↗

Reason for Recall

Product exhibits a decreased recovery of patient samples on the cobas e 801 module compared to the MODULAR ANALYTICS E 170, cobas e 411, 601, and 602 analyzers.

Distribution

US Distribution in states of: CA, FL and IN.

Lot / Code Info

Elecsys BRAHMS PCT: 07301715190 Lot number 204084

Root Cause

Software design

Action Taken

The firm, Roche, sent an "URGENT MEDICAL DEVICE REMOVAL" letter dated December 22, 2017 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: " Immediately discontinue use of Elecsys BRAHMS PCT (catalog number 07301715190) reagent lot number 204084 on the cobas e 801 module. " Discard affected product according to local guidelines. " Complete the attached fax form and fax it to 1-800-722-7222. " File this Urgent Medical Device Removal (UMDR) for future reference. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in this UMDR.