Zeiss brand OPMI Pentero C, Surgical microscope and accessories, Model: OPMI Pentero C, Product Numbers: 302583-9001-000 (PENTERO C), and 305961-0000-00 (Ceiling Mount for PENTERO C), Product Usage: A surgical microscope and accessories is an AC-powered device intended for use during surgery to provide a magnified view of the surgical field.

Company
Carl Zeiss Meditec Inc
Recall Initiated
April 18, 2013
Posted
May 31, 2013
Terminated
June 26, 2013
Recall Number
Z-1440-2013
Quantity
6 units
Firm Location
Dublin, CA

Reason for Recall

Carl Zeiss Meditec is conducting a field corrective action for its OPMI Pentero C surgical microscope system. During a routine inspection, one microscope installation was found to have 3 (three) broken screws out of 6 (six) in the lower section of the ceiling mount.

Distribution

US Nationwide Distribution.

Lot / Code Info

Serial Numbers: 6725000708, 6725000709, 6725000710, 6725000713, 6725100725, 6725120813

Root Cause

Under Investigation by firm

Action Taken

Carl Zeiss Meditec, Inc sent a Customer Notification letter to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter stated that firm's sales representative will be scheduling an on-site visit, informing them of the affected product and providing instructions on the recall. For question call 925-580-5371.