Siemens MEVATRON series, PRIMART (limited release)PRIMUS and/or ONCOR or ARTIST Linac Systems with BEAMVIEW capability. The intended use of the SIEMENS branded MEVATRON", ARTISTE", ONCOR" and PRIMUS" family of linear accelerator systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer.

This recall has been terminated (originally issued July 3, 2013).

Company
Siemens Medical Solutions Usa, Inc
Recall Initiated
May 29, 2013
Posted
July 3, 2013
Terminated
January 14, 2014
Recall Number
Z-1663-2013
Quantity
1956 customers
Firm Location
Martinez, CA
Official Source
View on FDA website ↗

Reason for Recall

The information regarding the FILM Mode calibration may have been considered difficult to interpret when translated. SIEMENS HEALTHCARE CR / RADIATION ONCOLOGY had become aware that this issue may have resulted in mistreatment of patients.

Distribution

Worldwide Distribution: USA Nationwide and countries of: Algeria, Angola, Argentina, Austria, Australia, Bahamas, Belgium, Bosnia Herzeg, Bulgaria, Brazil, Canada, Chile, Colombia, Croatia, Ecuador, Egypt, El Salvador, France, Germany, Greece, Hungary, India, Indonesia, Italy, Iran, Iraq, Ireland, Japan, Jordan, Kenya, DPR of Korea, Kosovo, Kuwait, Lebanon, Lithuania, Malaysia, Mexico, Nepal, Netherlands, New Zealand, Norway, Pakistan, Palestinian Ter, Peru, Philippines, P.R. China, Poland, Portugal, Romania, Saudi Arabia, Slovakia, Spain, Sri Lanka, South Africa, Sweden, Switzerlan, Taiwan, Thailand, Trinidad Tobago,Turkey, United Kingdom, Venezula, Vietnam, Rep of Yemen, and Zambia.

Lot / Code Info

All customers who use the BEAMVIEW for position verification imaging on Siemens MEVATRON series, PRIMART (limited release) PRIMUS and/or ONCOR or ARTIST Linac Systems.

Root Cause

Process change control

Action Taken

Siemens sent an "URGENT: MEDICAL DEVICE CORRECTION" letter and Update instruction TH022/13/S beginning on May 29, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. July 19, 2013 - TH023/13/S was distributed is now in progress. This is the FILM Mode Physics Primer Addendum.