RecallDepth

SIEMENS brand ARTISTE Linear Accelerator Linac systems with component: any photon unflat beam option (Multiple X feature) in combination with the IMRT or mARC option. A family of linear accelerator systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer. The ARTISTE/ONCOR systems are manufactured by: SIEMENS AG, MEDICAL SOLUTIONS, GERMANY and distributed SIEMENS Healthcare IM CR RO, CR Radiation Oncology, Martinez, CA

This recall has been terminated (originally issued December 3, 2013).

Company
Siemens Medical Solutions Usa, Inc
Recall Initiated
October 23, 2013
Posted
December 3, 2013
Terminated
January 22, 2014
Recall Number
Z-0438-2014
Quantity
26 units
Firm Location
Martinez, CA
Official Source
View on FDA website ↗

Reason for Recall

A safety risk regarding a potentially existing dark current radiation phenomenon on the Linear Accelerator (LINAC) in combination with IMRT or mARC (rotational IMRT) treatments using unflat (flattening filter free) beams. In case there are many long periods of field shape changes in a treatment plan, the dose due to dark current radiation may sum up to a significant value.

Distribution

Worldwide Distribution: USA (nationwide) including states of: FL, NC, NY, PA, TN, UT, and WI; and countries of: Angola, Brazil, Canada, P.R. China, Germany, India, Japan, Republic Korea, Poland, and Thailand.

Lot / Code Info

Serial Numbers:  5364, 5419, 5640, 5737, 5823, 5830, 5166, 5342, 5416, 5444, 5479, 5485, 5628, 5745, 5746, 5762, 5763, 5764, 5772, 5781, 5783, 5797, 5815, 5819, 5820, 5822, 5829, 5837, 5839, 5888, 5896, 5911, 5937

Root Cause

Device Design

Action Taken

Siemens sent an Urgent Medical Device Correction letter dated October 23, 2013 including the release of update instruction TH024/13/S to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer.

More recalls by Siemens Medical Solutions Usa, Inc