Medtronic Octopus Nuvo Tissue Stabilizer, model TSMICS1. The Octopus Nuvo Tissue Stabilizer is a disposable tissue stabilization system consisting of a rigid shaft, detachable tissue stabilizer headlink and a flexible articulating arm. The device is assembled within the thoracic cavity. Sterilized using ethylene oxide, non pyrogenic, disposable, single use only. The product is intended to stabilize and minimize the motion of selected areas of the beating heart during minimally invasive cardiac surgery procedures under direct visualization through a thoracotomy.

This recall has been terminated (originally issued November 1, 2010).

Company
Medtronic Perfusion Systems
Recall Initiated
September 13, 2010
Posted
November 1, 2010
Terminated
April 16, 2012
Recall Number
Z-0134-2011
Quantity
571 units
Firm Location
Brooklyn Park, MN
Official Source
View on FDA website ↗

Reason for Recall

Medtronic has received reports regarding the Octopus Nuvo Tissue Stabilizer stating that during use, the collet (a component of the device that stabilizes the head link) fractured, thereby causing immediate separation of the head link from the shaft of the device. The resulting potential hazards are that fragments of the collet could fall into the patient's chest cavity and/or damage the heart ti

Distribution

within the US to: AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, MA, MI, MN, MS, NJ, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA, WI and Washington D.C. OUS to include: Belgium, Canada, Germany, Italy, Poland, Switzerland, United Kingdom.

Lot / Code Info

Lot #: 201002P030, 2010030795, 2010032195, 2010032196, 2010032690, 2010033377, 2010040102, 2010040103, 2010040673, 2010041001, 2010041500, 2010061786, 2010062722, 2010071111, 2010081105, 2010081987.

Root Cause

Component design/selection

Action Taken

Medtronic sent an "Urgent Medical Device Recall Notice" dated September 14, 2010, to Risk Managers at each affected account. The letter described the issue, identified affected lot numbers, asked for immediate discontinued use of the device, quarantine all unused devices and to return unused devices to Medtronic CardioVascular, 7611 Northland Drive, Brooklyn Park, MN 55428 and requested a recall certificate be faxed back to Medtronic indicated acknowledgement of the Recall Notice and quantity to be returned. If you need additional information, please contact Medtronic Cardiovascular Lifeline Technical Services at 877-526-7890, or your local Medtronic sales representative.