ACUSON S2000 Automated Breast Volume Scanner (ABVS); Model number 10042541 Distributed by Siemens Medical Solutions USA Inc. business Unit: Ultrasound. The S2000 Automated Breast Volume Scanner (ABVS) Ultrasound System for imaging of the breast allows volumetric acquisition of the breast image.

This recall has been terminated (originally issued March 9, 2013).

Company
Siemens Medical Solutions Usa, Inc.
Recall Initiated
January 18, 2013
Posted
March 9, 2013
Terminated
July 16, 2013
Recall Number
Z-0930-2013
Quantity
214 affected devices
Firm Location
Mountain View, CA
Official Source
View on FDA website ↗

Reason for Recall

On some ACUSON S2000 ABVS systems, the inside of the transducer pod may contain sharp edges.

Distribution

Worldwide Distribution - USA including WV and OH; Internationally to Australia, Austria, Bahrain, Belgium, Brazil, China (P.R.) Estonia, France, Germany, Greece, India, Indonesia, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Macedonia, Malaysia, Mexico, Netherlands, Peru, Philippines, Poland, Portugal, Qatar, Rep of Korea, Romania, Sweden, Taiwan, Thailand, Turkey, Unite Arab Emirates, United Kingdom and Venezuela.

Lot / Code Info

Model number 10042541 - AVBS module.  Note: the S2000 system part number is 10041461, however only those systems that include the ABVS module, part number 10042541 are affected. Manufactured prior to 10/30/2012.

Root Cause

Process design

Action Taken

Siemens sent a Customer Safety Advisory notification letter to Siemens Service Personnel on January 18, 2013 and which would be provided to all affected sites during their visit to correct the problem. The notification identified the product, the problem and the action to be taken. The local Customer Service Engineer will perform the test at their site. If the system passes the test, no further action is required. If the ABVS pod fails the inspection, the Service Engineer will correct the sharp edge. For questions regarding this recall call 650-694-5398.