ACUSON SC2000 Ultrasound System. The firm name on the label is Siemens Medical Solutions USA, Inc., Buffalo Grove, IL.

Company
Siemens Medical Solutions Usa, Inc.
Recall Initiated
March 5, 2018
Posted
March 26, 2018
Terminated
February 26, 2021
Recall Number
Z-1200-2018
Quantity
229 systems
Firm Location
Mountain View, CA

Reason for Recall

The ECG signal may flatline due to electromagnetic interference during the use of electrosurgical equipment.

Distribution

Distribution was made to medical facilities in AZ, CA, CT, FL, IL, IN, KS, KY, MI, MN, MO, NJ, NY, NC, OH, PA, TX, VA, and WA. There was no military/government distribution. Foreign distribution was made to Argentina, Australia, Austria, Belgium, Bosnia Herzeg, Colombia, Croatia, Denmark, France, Germany, India, Ireland, Israel, Italy, Japan, Mexico, New Zealand, Philippines, Poland, Republic Korea, Romania, Singapore, Slovakia, Spain, Sweden, Taiwan, Thailand, Turkey, UAE, and Vietnam.

Lot / Code Info

Software versions VB20A, VB20B, or VB20C that have a common physio module (CPM) (known by customers as the ECG (physio) hardware) installed at revision 4

Root Cause

Device Design

Action Taken

Undated recall letters were issued via email (if they had an email address for the customer) or via certified mail (if no email address was available) on 3/5/2018.