NxStage Dialysate Sack (SAK) with Concentrate for Use with PureFlow SL REF: SAK-407 Lactate 40, Potassium 1, Sodium 140, Calcium 3, Magnesium 1, Chloride 105, Glucose (mg/dL) 100 Batch Size (liters) 50

Company
Nxstage Medical, Inc.
Recall Initiated
May 14, 2014
Posted
June 24, 2014
Terminated
March 13, 2017
Recall Number
Z-1861-2014
Quantity
91 cases
Firm Location
Lawrence, MA

Reason for Recall

SAK Dialysate Concentrate contain aluminum levels which exceed internal product specification

Distribution

Worldwide Distribution - USA (nationwide) and Internationally to France, Netherlands, Spain, Sweden, and the UK.

Lot / Code Info

Lot code: 40179074

Root Cause

Material/Component Contamination

Action Taken

NxStage issued Initial recall letter dated 5/14/14 to dialysis centers. An error was noted and an updated letter dated 5/15/14 was resent to dialysis centers. The updated 5/15/14 letter was emailed to all current active PureFlow SL patients who had provided an email address followed by UPS next business day letter sent to all current active PureFlow SL patients. Accounts are requested to examine inventory and remove product. Contact NxStage Customer Service to arrange for return of all affected product and for replacement product to be sent. If you have any questions or comments, please feel free to contact NxStage Customer Service at 1-866-NXSTAGE (1-866-697-8243).