Aesculap proSA Adjustment Disc Size Large (L) The Miethke proSA Adjustment shunt system is intended to shut cerebrospinal fluid from the lateral ventricles of the brain into the peritoneum.

Company
Aesculap Inc
Recall Initiated
December 23, 2013
Posted
February 19, 2014
Terminated
October 1, 2014
Recall Number
Z-0989-2014
Quantity
79
Firm Location
Center Valley, PA

Reason for Recall

The accuracy is out of specification for the Aesculap proSA Adjustment Disc Size L.

Distribution

Worldwide Distribution - US Distribution including the states of AZ, MD, MA, WI, MI, MN, DE, KS, and FL., and the countries of Australia, Germany, Great Britain, Norway and Japan.

Lot / Code Info

4 lots: 4505268906, 4505336973, 4505408485, 4505238953

Root Cause

Employee error

Action Taken

Aesculap sent an Important Correction and Removal Notification dated December 23, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers.. Yom"Aesculap Sales Representative will contact you to schedule a time to remove the FV795T from Set FV792T. Please use FV790T in the absence of the FV795T to ensure the correct adjustment has been made. Please complete the attached Inventory Sheet which is necessary to comply with FDA regulations. When completing the inventory sheet, please fill in the quantity being returned. If you cannot locate the product, please provide an explanation as to why the inventory will not be returned (discarded, etc.). AIC (USA) appreciates your cooperation on this matter and apologizes for the inconvenience this may cause. Thank you for your patience and continued support of this product. Please call (610) 984-9265 or (610) 984-9414 with any questions.

More recalls by Aesculap Inc

View all recalls by this company →