TxRx Knee 15 Coil 3T. For use in conjunction with magnetic resonance scanners to produce diagnostic images of the knee that can be interpreted by a trained physician.

This recall has been terminated (originally issued September 6, 2012).

Company
Quality Electrodynamics Llc
Recall Initiated
July 27, 2012
Posted
September 6, 2012
Terminated
February 19, 2013
Recall Number
Z-2314-2012
Quantity
17 units
Firm Location
Mayfield Village, OH
Official Source
View on FDA website ↗

Reason for Recall

On July 27, 2012, it was reported to the firm that the Siemens 3T 15ch Knee Coil was not functional because the coil identification code for the device was incorrect.

Distribution

Worldwide distribution: USA (nationwide): Tennessee only and country of: Germany

Lot / Code Info

QED part numbers: Q7000057/QED000057/QSV000057 Siemens part numbers: 10606829 / 10606614  Serial #s: 1002-1012, 1016-1020, and 1022.

Root Cause

Error in labeling

Action Taken

Since all of the devices were located at various Siemens distribution locations, all communications were internal to Siemens concerning bringing the coils from one warehouse to another.