TxRx Knee 15 Coil 1.5T QED part number: Q7000056 Siemens Model # 10606828

This recall has been terminated (originally issued August 20, 2019).

Company
Quality Electrodynamics Llc
Recall Initiated
August 20, 2019
Terminated
May 26, 2020
Recall Number
Z-0036-2020
Quantity
36 units
Firm Location
Mayfield Village, OH
Official Source
View on FDA website ↗

Reason for Recall

Unintended ferrous material may be present within the pushbutton assembly of Siemens Knee and Siemens Extremity Coils.

Distribution

Worldwide distribution - US in the states: NC OUS: Germany

Lot / Code Info

Serial # 1694 1841 1887 1933 1998 3195 3291 3292 3293 3294 3298 3299 3300 3301 3302 3303 3304 3305 3306 3307 3308 3309 3310 3311 3312 3313 3314 3315 3316 3317 3318 ¿1137 ¿1361 ¿1378 ¿1628  ¿2447

Root Cause

Nonconforming Material/Component

Action Taken

The firm sent a notification letter to their customers who that have received the affected device. Customers were instructed to. 1. Immediately examine your inventory and quarantine product subject to recall. 2. if you have further distributed this product to MRI end users, please identify your affected customers and notify them. 3. Please ensure they understand to stop use of the coil for scanning and to notify you immediately.