SUN|DASH RADIUS 252 tanning system; 52 lamps, Ultraviolet Radiation. The device is intended to be used for tanning of the human skin.

This recall has been terminated (originally issued September 7, 2012).

Company
Jk Products & Services, Inc
Recall Initiated
July 8, 2005
Posted
September 7, 2012
Terminated
January 13, 2017
Recall Number
Z-2334-2012
Quantity
1948 units
Firm Location
Jonesboro, AR
Official Source
View on FDA website ↗

Reason for Recall

The firm has tracked a systematic failure in the door cord on the Sun|Dash Radius 252 product line. In some instances when the door cord is installed with little or no slack outside the upper decor piece, the cord has been known to fail causing a short in the wiring at the pivot point in the cord.

Distribution

Worldwide Distribution - USA (nationwide) and Canada

Lot / Code Info

PN 8000809705; Serial numbers 248-1298

Root Cause

Equipment maintenance

Action Taken

JK Global Services sent out Mandatory Inspection/Modification letters to their Distribution Service Teams dated 07/08/2005, 01/12/2006, 09/15/2010, 10/04/2010 and 03/01/2012. A letter dated September 15, 2010, was sent to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were asked to contact a JK representative or their Distributor as soon as possible to schedule an inspection to the affected product in their possession. For any questions customers were instructed to call 1-800-445-0624 or send an e-mail to support@jkamerica.com.