RecallDepth

Oridion CO2 sampling lines and water traps under the label: MicroStreamC02 Sampling Lines: FilterLine Set Adult/Pediatric -Part No: XS04620-01; FilterLine Set Adult/Pediatric (100 unit boxes)-Part No. 010579; FilterLine H Set Adult/Pediatric -Part No. XS04624 ; FilterLine H Set Adult/Pediatric (100 unit boxes) Part No: 010580; FilterLine H Set Infant/Neonatal-Part No. 006324; FilterLine Set Adult/Pediatric Long Part No. 007768; FilterLine H Set Adult/Pediatric Long Part No. 007737; FilterLine H Set Infant/Neonatal Long Part No. 007738 Smart CapnoLine Plus (O2 connector) Part No. 009818 Smart CapnoLine Plus (O2 connector) (Special 100 unit boxes) - Part No.; 010209; Smart CapnoLine Plus Long (O2 connector) Part No. 010340; Smart CapnoLine Plus Long (O2 connector) (Special 100 unit boxes) Part No: 010339; Smart CapnoLine Plus O2 (O2 tubing) Part No. 009822 Smart CapnoLine Plus O2 (O2 tubing) (Special 100 unit boxes) Part No. 010210; Smart CapnoLine Plus O2 Long (O2 tubing) Part No. 009826; Smart CapnoLine Plus O2 Long (O2 tubing) (Special 100 unit boxes) , Part No. 010341; Smart CapnoLine Pediatric Part No. 007266 Smart CapnoLine O2 Pediatric (O2 tubing) Part No. 07269; Smart CapnoLine O2 Pediatric Long (O2 tubing) Part No.007743; Smart CapnoLine H Plus O2 (O2 tubing) Part No. 010433; Smart CapnoLine H O2 Adult (O2 tubing) Part No. 010478; Smart CapnoLine H O2 Intermediate (O2 tubing) Part No; 010475; Smart CapnoLine H Pediatric (O2 connector) Part No. 010581; Smart CapnoLine H O2 Pediatric (O2 tubing) Part No. 010582 O2 CO2 Nasal FilterLine Adult (O2 connector) Part No. 010207; O2 CO2 Nasal FilterLine Adult (O2 tubing) Part No. 006912; O2 CO2 Nasal FilterLine Adult Long (O2 connector) Part No. 010342; O2 CO2 Nasal FilterLine Adult Long (O2 tubing) Part No. 007739; O2 CO2 Nasal FilterLine Pediatric (O2 tubing) Pat No. 006913; O2 CO2 Nasal FilterLine Long (O2 tubing) Part No 007740; NIV Line Adult Part No. 008174; NIV Line Pediatric Part No. 008175; Nasal FilterLine Infant/Neonatal Part No. XS0447; CapnoLine H Adult Part No. 008177; CapnoLine H Pediatric Part No. 008178; CapnoLine H Infant/Neonatal Part No. 008179; CapnoLine H O2 Adult (O2 tubing) Part No. 008180; CapnoLine H O2 Pediatric (O2 tubing) Part No. 008181; CapnoLine H O2 Infant/Neonatal (O2 tubing) Part No. 012111; VitaLine H Set Adult/Pediatric Part No. 010787; VitaLine H Set Infant/Neonatal Part No. 010807; FilterLine XL Part No. 006325; Smart CapnoLine Guardian Part No. 012528; Smart CapnoLine Guardian O2 Part No. 012529; Smart CapnoLine Guardian O2 Long Part No. 012530

This recall has been terminated (originally issued March 29, 2012).

Company
Oridion Medical
Recall Initiated
February 16, 2012
Posted
March 29, 2012
Terminated
September 13, 2013
Recall Number
Z-1332-2012
Quantity
2 million
Official Source
View on FDA website ↗

Reason for Recall

CO2 sampling lines may have paper and plastic particulate present in the packaged units at levels that may exceed visual specification.

Distribution

Nationwide Distribution

Lot / Code Info

All lots

Root Cause

Packaging process control

Action Taken

Oridion notified Hospitals, Clinics, EMS with an "Important Communication" letter to clinicians dated February 12, 2012. The letter was intended to make US clinicians aware of an important issue relating to certain Oridion CO2 sampling lines, which may not fully meet quality requirements. In a few cases, paper and plastic particulate may be present in the packaged units at levels that exceed the visual specification. Recommended actions were included in the letter. Questions were directed to 1(781)-972-1252.

More recalls by Oridion Medical

View all recalls by this company →