RecallDepth

Bausch & Lomb SoFlex SE Foldable Intraocular Lens. UPC 4001404, Rx Only, STERILE. Manufactured by: Bausch & Lomb Incorporated, 21 Park Place Boulevard, Clearwater, FL 33759. Intraocular lens indicated for primary implantation for visual correction in adult patients where the cataractous lens has been removed by phacoemulsification. The lens is intended for placement in the capsular bag.

This recall has been terminated (originally issued January 11, 2010).

Company
Bausch & Lomb Inc
Recall Initiated
November 19, 2009
Posted
January 11, 2010
Terminated
July 22, 2010
Recall Number
Z-0591-2010
Quantity
85 lenses
Firm Location
Rochester, NY
Official Source
View on FDA website ↗

Reason for Recall

Bausch & Lomb Incorporated, Rochester, NY is recalling their Silicone L161SE Intraocular Lens Lot 4916928 (23.5D) after receiving three customer complaints for what appears to be opacification of the intraocular lens. The opacification occurred within one week of implantation. In each reported case the surgeon explanted the lens.

Distribution

Lenses were distributed to one distributor and 35 retailers. Worldwide Distribution -- Distributed to Canada, Taiwan and throughout the United States.

Lot / Code Info

Model Number: LI61SE, Lot Number: 4916928.

Root Cause

Other

Action Taken

Bausch & Lomb, Inc. notified consignees via phone beginning November 19, 2009 and by follow-up letter dated November 23, 2009. Titled "Urgent - Medical Device Recall", the letter instructed users to discontinue implantation of the affected product and return any unused lenses along with the enclosed acknowledgement form to the firm. For further information, contact Bausch & Lomb at 1-585-338-6612.