Philips BV300 Mobile X-Ray System

This recall has been terminated (originally issued February 1, 2003).

Company
Philips Medical Systems
Recall Initiated
September 27, 2002
Posted
February 1, 2003
Terminated
January 12, 2005
Recall Number
Z-0526-03
Quantity
229 devices
Firm Location
Bothell, WA
Official Source
View on FDA website ↗

Reason for Recall

The C-bow support can bend and break allowing C-arm to fall from support

Distribution

The firm distributed to hospitals, x-ray centers and medical centers throughout the U.S.

Lot / Code Info

Part numbers 9896 000 06801 and 9806 000 06802 with serial numbers: DBXXXX - all DBXXXX - all DDXXXX - all DEXXXX - all DG0001-DG0129 DH0001-DH0033 DJ0001 - DJ0152 DK0001 - DK 0076

Root Cause

Other

Action Taken

The firm issued a letter dated 9/27/02 to their customers. The letter advised of the problem and that a service representative will visit to inspect unit and correct.