Philips BV300 Mobile X-Ray System

Company
Philips Medical Systems
Recall Initiated
September 27, 2002
Posted
February 1, 2003
Terminated
January 12, 2005
Recall Number
Z-0526-03
Quantity
229 devices
Firm Location
Bothell, WA

Reason for Recall

The C-bow support can bend and break allowing C-arm to fall from support

Distribution

The firm distributed to hospitals, x-ray centers and medical centers throughout the U.S.

Lot / Code Info

Part numbers 9896 000 06801 and 9806 000 06802 with serial numbers: DBXXXX - all DBXXXX - all DDXXXX - all DEXXXX - all DG0001-DG0129 DH0001-DH0033 DJ0001 - DJ0152 DK0001 - DK 0076

Root Cause

Other

Action Taken

The firm issued a letter dated 9/27/02 to their customers. The letter advised of the problem and that a service representative will visit to inspect unit and correct.