Gyroscan 1.5T Intera

This recall has been terminated (originally issued September 12, 2003).

Company
Philips Medical Systems
Recall Initiated
May 2, 2002
Posted
September 12, 2003
Terminated
April 11, 2005
Recall Number
Z-1201-03
Quantity
see above
Firm Location
Bothell, WA
Official Source
View on FDA website ↗

Reason for Recall

Patient burns related to high SAR levels and the combined use of Synergy Body Coil and synergy Flex-M coil

Distribution

The firm distributed units to hospitals and medical centers nationwide.

Lot / Code Info

Serial numbers provided above

Root Cause

Other

Action Taken

The firm issued a Field Change Order to their field personnel instructing them to add new labeling to the User Manual. 05/02/2002 A letter was issued to customers.